The DP-301 is a single channel AC/DC differential pre-amplifier highly suited for EEG, EKG and extracellular recording.
|
High common mode rejection is very important in minimizing electrical interference. The high input resistance of the DP-301 (1012 ohm typical) ensures that the instrument's high CMR will not be degraded by differences in source impedance at the input. This is important in extracellular recording, where the difference in resistance of the recording and indifferent electrode is often large.
Extracellular action potentials are typically measured in microvolts but are usually accompanied by much larger DC electrode voltages. The DP-301 is designed to amplify these microvolt level events with as much as ± 3 VDC at the input.
Gain selections of x100, x1000 and x10000. Microvolt signals are sufficiently amplified for computer data acquisition and recorder inputs. Noise is very low, typically 10 µV p-p at 1 Hz to 10 kHz bandwidth. Reducing the bandwidth with the low pass and high pass filters further lowers noise.
The internal Cal signal provides a convenient check of amplifier gain and operation.
The single channel DP-301 is powered with 4 standard 9 V transistor batteries yielding approximately 500 hours of normal operation. The small size of the DP-301 makes it possible to be placed at the measurement site or inside a Faraday cage. End of battery life is indicated by the Low Bat light.
The standard IC-2S input cables (2 per channel) which are not terminated at the outboard end.
Optionally available are cables with 2 mm pin and alligator clip (p/n IC-4M).
IC-4M Cable shown to right